FLUO-CTRLMedical Imaging

Fluoroscopy Control Room

Standard Area13
Ceiling Height2700mm

Briefing

Description
The Fluoroscopy Control Room provides the space and equipment for remote control of fluoroscopy equipment and amenities for the review of images and reporting. Direct visibility of the patient through an observation window and direct access to the imaging room is required.
Hours of Operation
Up to 24 hours
Occupancy
1-2 staff
Additional Considerations
~ Cable ducts for services between the imaging device and the control area will be required and must be coordinated with the MME vendor. Additional building services connections and outlets are to be coordinated to suit equipment requirements. ~ All shielding on this Standard Component is indicative. Radiation shielding to suit the overall department layout, anticipated workload and the equipment procured is to be assessed and advised by a qualified Radiation Consultant to meet relevant legislation and local radiation safety policies. ~ An ‘X-ray in Use’ light system is required with signs at each entry to the imaging room that are illuminated when the imaging device is on and turned off when it stops. The exact wording required on illuminated signs will be determined by relevant legislation and local radiation safety policies. Project teams must confirm all radiation safety signage requirements with their jurisdiction’s regulatory agency. ~ The size, quantity and configuration of computer display screens is to be finalised based on the specific image display and processing task requirements (e.g. number of application windows to be viewed concurrently) and ergonomic considerations. Providing flexibility of monitor positioning is vital to ensure adjustments are possible to meet the ergonomic requirements of different users. ~ Provision of height adjustable workstations or fixed joinery is to be in line with service requirements and local WHS policies. ~ Hours of operation depend on the service in which this room is located. Ambulatory medical imaging services may only operate during business hours or extended business hours, some small hospital services may have an on-call, after-hours service. Large units are likely to provide a 24-hour service, especially where a satellite service is provided with an emergency department. ~ Requirement for fixed and/or mobile duress to be confirmed to suit service requirements. ~ An operational model for responding to calls from staff/patients will be needed to support the inclusion of a nurse call system and is to be determined based on service requirements. ~ Provision of an intercom for staff communication within the department is dependent on service requirements. This function may be provided as part of the telephone system. ~ The need for and extent of emergency power and UPS must be reviewed to suit site specific requirements and confirmed based on risk assessment considering the impact of a power outage on patient care/safety.

Performance Requirements

HVAC

Air Conditioning (General)

Lighting

GeneralDimmable

Nurse Call

Buttons / Handset

Items

CodeDescriptionQtyPriorityCategoryComment
AFDPR-001.01DOOR PROTECTION: kickplate, to 300H1StandardInternal - Door 1to control room side
AFDPR-001.01DOOR PROTECTION: kickplate, to 300H1StandardInternal - Door 2to control room side
AFDPR-006.01DOOR PROTECTION: plate, to 900H1StandardInternal - Door 2[DWPR-005] to corridor side, extent dependent on movement of beds and mobile equipment in adjacent area; to be coordinated with corridor wall protection
AFDPR-006.01DOOR PROTECTION: plate, to 900H1StandardInternal - Door 1to imaging room side
AHWCV-001WINDOW COVERING: blind, roller, single1StandardInternal - Window 1
DOHI-013.01DOOR: hinged, 1 leaf, 900 clear opening, solid, standard vision panel1StandardInternal - Door 2optional - staff entry from corridor
DOHI-013.05DOOR: hinged, 1 leaf, 900 clear opening, solid, shielded, standard vision panel, lead glass1StandardInternal - Door 1to imaging room
WIFX-008.03WINDOW: fixed, internal, single glazed, sill at 1050H, shielded, lead glass1StandardInternal - Window 1for observation from control room