Back to Components
CTPRRadiation Oncology

CT Planning Room

Standard Area45
Ceiling Height3000mm

Briefing

Description
The CT Planning Room provides space and facilities for Computed Tomography (CT) scanning procedures as part of the planning of radiation therapy treatment for patients. A collocated Control Room with observation window is required for the remote operation of equipment.
Hours of Operation
8am to 6pm, daily (dependent on unit operating hours)
Occupancy
1 patient; up to 4 staff assisting
Additional Considerations
~ An Equipment Room may be required and will need to be in close proximity due to acceptable cable lengths. ~ Cable ducts between the Scanning Room, Control Room and Equipment Room will be required. Equipment is to be installed to manufacturer's specifications with sufficient clearances for maintenance access. ~ The CT Planning Room must be adequately shielded. Radiation shielding is to be as advised and confirmed by a Radiation Consultant to suit the equipment procured. 'X-ray in Use' lights are required outside each entry door to be illuminated when scanning begins. ~ Interlock system to doors recommended to prevent machine from running if door open and to stop the machine upon entry if in use. ~ Additional medical gases may be provided (including additional oxygen, additional suction, nitrous oxide and scavenge) depending on clinical requirements, such as the provision of anaesthetic services. ~ Dimmable lights are to be provided and lights are to be placed to avoid glare to patients. ~ Back lit ceiling artwork may be installed for patient comfort and distraction. ~ Where possible, it is preferred that ductwork in the ceiling is designed to the perimeter of room to allow for future flexibility to provide ceiling mounted equipment. ~ The need for and extent of emergency and UPS power must be reviewed to suit site specific requirements and confirmed based on risk assessment considering the impact of a power outage on patient care/safety. ~ Unistrut ceiling able to carry moving loads may be required. ~ A mix of respiratory gating, surface guidance, motion management and/or patient position technologies may be provided depending on clinical/service requirements and will be coordinated to support the technologies provided in the Linear Accelerator Treatment Room. The provision of these technologies is to be confirmed by the project and may have additional design requirements including additional power/data, water cooling, heat load, possible increased set down, increased number of trenches/conduits, space requirements for computer cabinets in control room and additional penetrations through the suspended ceiling. Requirements to be confirmed with vendors.

Performance Requirements

HVAC

Air Conditioning (General)

Lighting

DimmableColour Corrected

Electrical

Body Protected

Nurse Call

Buttons / Handset

Items

CodeDescriptionQtyPriorityCategoryComment
AFDPR-006.01DOOR PROTECTION: plate, to 900H1StandardInternal - Door 2[DWPR-005] to control room side
AFDPR-006.01DOOR PROTECTION: plate, to 900H3StandardInternal - Door 1[DWPR-005] to imaging room side
AFDPR-006.01DOOR PROTECTION: plate, to 900H1StandardInternal - Door 2[DWPR-005] to imaging room side
AFDPR-006.01DOOR PROTECTION: plate, to 900H1StandardInternal - Door 1[DWPR-005] to corridor side, extent dependent on movement of beds and mobile equipment in adjacent area; to be coordinated with corridor wall protection
DOHI-003.02DOOR: hinged, 1 leaf, 900 clear opening, solid, shielded1StandardInternal - Door 2[DWSC-005] to control room
DOHI-204.02DOOR: hinged, 2 leaves, 1800 clear opening, solid, shielded1StandardInternal - Door 1[DWSC-034] to corridor
WIFX-007.03WINDOW: fixed, internal, single glazed, sill at 900H, shielded, lead glass1StandardInternal - Window 1[DWOB-009] safety glass, to control room